Carci Fazendo por você
Phone contact Carci (55 11) 3346-2100

Shortwave Diathermy

  • Shortwave Diathermy - Shortwave Diathermy - Diatermed II
  • Shortwave Diathermy - Shortwave Diathermy - Diatermed II
  • Watch the video Shortwave Diathermy - Diatermed II

Shortwave Diathermy - Diatermed II

4022

Continuous and pulsed (burst) mode
Capacitive Electrode boards and Schiliephake
Potency: 240W continuous mode / 300W pulsed mode
Pulse width (duration): 400uS
Pulse Frequency: 45, 70, 85, 115, 145, 175, 200, 230, 300 and 400Hz
With 3 protection systems: operation mode changes, patient protection key and momentary power failure.
Schliephake electrodes with adjustable graduation distance (electrode-skin): 0.0 cm to 2.5 cm for better heat distribution.



Operation mode: continuous and burst
Time switch: 1 to 30 minutes (step 5 minutes) with acoustic signalization and exit circuit off.
Board dimensions: 12x17cm
Schiliephake disc dimension: 13 or 8,5cm ø

Continuous Mode
Potency (50 Ohms charged) board electrodes: 240W± 20%
Potency (50 Ohms charged) schiliephake: 200W± 20%

Burst Mode
Pulse width (duration): 0,4ms. (± 5 %)
Pulse frequency: 45, 75, 85, 115, 145, 175, 200, 230, 300, 400 Hz (± 5%)
Potency (50 Ohms charged): 300W (peak) ± 30%
Operation frequency: 27,12 Mhz ± 1%

Tuning: adjustment with indicator in leads bar
Intensity control: sensible and linear adjustment
Mains Power: 127 /220 Vac 50/60Hz
Fuse: 5A to 127V ~ (250V-20AG)
3A to 220V ~ (250V-20AG)
Maximum consumption: 500 VA
Dimensions: 49x 74 x 30cm (WxHxD)
Weight: 33 Kg
Safety Tests: IEC 60601-1, IEC60601-2, IEC 0601-2-3

 

Estudio Bigas Desenvolvimento de sites